TERUMO INFUSION PUMP

Product name
TERUFUSION Infusion Pump Type LF
Model
TE-LF600
Flow rate setting range
• Setting range
3.00 to 300.00 mL/h (when setting is 20 drops/mL)
1.00 to 100.00 mL/h (when setting is 60 drops/mL)
• Step
1.00 mL/h step*
*: Step can be used when ow rate is set by the Dial. (It will be 0.00 mL/h when the power is turned on.)
Flow rate accuracy
Within ±10%
(The hourly precision after one hour has passed since the start of solution delivery, in the case where water or a physiological saline solution is used at a constant ow rate of minimum ow rate or more (ambient temperature: 23±2°C) according to IEC 60601-2-24 standard.)
Within ±5%
(The hourly precision after one hour has passed since the start of solution delivery, in the case where water or a physiological saline solution is used with TERUFUSION Solution Administration Set (T_Grav.) at minimum ow rate (20 drops/mL: 3.00 mL/h or 60 drops/mL: 1.00 mL/h) or middle ow rate (25.00 mL/h) (ambient temperature: 23±2°C, drop between infusion bag and pump: 50 cm, and drop between pump main body and puncture site: None) according to IEC 60601-2-24 standard.)
Occlusion detection pressure
• Upper occlusion detection pressure -100 to -30 kPa
• Lower occlusion detection pressure
30 to 120 kPa(set value)
The lower occlusion detection pressure can be set in 10 levels.
Purge ow rate
Approx.300 mL/h or more (depending on the infusion set to be used)
Alarms
Upper Occlusion alarm, Lower Occlusion alarm, Air-in-line alarm, Door alarm, Flow Rate Abnormality alarm, Free Flow alarm, Line Empty alarm, Battery alarm, Re-alarm, Start Reminder, No Flow Rate alarm, No VTBI alarm, Flow Rate/VTBI Volume Judgment alarm, Completion alarm
Safety functions
Occlusion detection pressure select function, Air-in-line alarm sensitivity select function, Tube clamp function, Flow rate range setting function A (Soft limits), Flow rate range setting function B (Hard limits), Key lock function
Use conditions
Ambient temperature: 5 to 40°C*
Relative humidity: 20 to 90%RH (non-condensing)
*: Flow Rate Abnormality alarm may be issued under approx. <10°C at a ow rate of approx. 50mL/h or more when using TERUFUSION Solution Administration Set (T_Grav.). Storage conditions Ambient temperature: -20 to 45°C Relative humidity: 10 to 95%RH (non-condensing) Power supply AC100 to 240V, 50 to 60Hz Internal battery (lithium ion battery) • Continuous use period: Approx. 5 hours (applies when solution is delivered continuously at 25 mL/h in ambient temperature of 25°C with new fully charged battery) • Charging time: ≥ 8 hours (when charged with AC power supply with the power turned off) Power consumption 28VA Classi cation Class I equipment and internally powered equipment, type CF applied part, Continuous Operation, IP22 (Vertically dripping water shall have no harmful effect when the enclosure is tilted at an angle up to 15° from its normal position.) Dimensions Without external communication (RS-232C): 253 mm (W)*1 x 120 mm (H)*1 x 102 mm (D)*2 With external communication (RS-232C): 253 mm (W)*1 x 120 mm (H)*1 x 121 mm (D)*2 *1: Excluding protrusions, *2: Excluding protrusions, pole clamp and moving range

Description

Product name
TERUFUSION Infusion Pump Type LF
Model
TE-LF600
Flow rate setting range
• Setting range
3.00 to 300.00 mL/h (when setting is 20 drops/mL)
1.00 to 100.00 mL/h (when setting is 60 drops/mL)
• Step
1.00 mL/h step*
*: Step can be used when ow rate is set by the Dial. (It will be 0.00 mL/h when the power is turned on.)
Flow rate accuracy
Within ±10%
(The hourly precision after one hour has passed since the start of solution delivery, in the case where water or a physiological saline solution is used at a constant ow rate of minimum ow rate or more (ambient temperature: 23±2°C) according to IEC 60601-2-24 standard.)
Within ±5%
(The hourly precision after one hour has passed since the start of solution delivery, in the case where water or a physiological saline solution is used with TERUFUSION Solution Administration Set (T_Grav.) at minimum ow rate (20 drops/mL: 3.00 mL/h or 60 drops/mL: 1.00 mL/h) or middle ow rate (25.00 mL/h) (ambient temperature: 23±2°C, drop between infusion bag and pump: 50 cm, and drop between pump main body and puncture site: None) according to IEC 60601-2-24 standard.)
Occlusion detection pressure
• Upper occlusion detection pressure -100 to -30 kPa
• Lower occlusion detection pressure
30 to 120 kPa(set value)
The lower occlusion detection pressure can be set in 10 levels.
Purge ow rate
Approx.300 mL/h or more (depending on the infusion set to be used)
Alarms
Upper Occlusion alarm, Lower Occlusion alarm, Air-in-line alarm, Door alarm, Flow Rate Abnormality alarm, Free Flow alarm, Line Empty alarm, Battery alarm, Re-alarm, Start Reminder, No Flow Rate alarm, No VTBI alarm, Flow Rate/VTBI Volume Judgment alarm, Completion alarm
Safety functions
Occlusion detection pressure select function, Air-in-line alarm sensitivity select function, Tube clamp function, Flow rate range setting function A (Soft limits), Flow rate range setting function B (Hard limits), Key lock function
Use conditions
Ambient temperature: 5 to 40°C*
Relative humidity: 20 to 90%RH (non-condensing)
*: Flow Rate Abnormality alarm may be issued under approx. <10°C at a ow rate of approx. 50mL/h or more when using TERUFUSION Solution Administration Set (T_Grav.). Storage conditions Ambient temperature: -20 to 45°C Relative humidity: 10 to 95%RH (non-condensing) Power supply AC100 to 240V, 50 to 60Hz Internal battery (lithium ion battery) • Continuous use period: Approx. 5 hours (applies when solution is delivered continuously at 25 mL/h in ambient temperature of 25°C with new fully charged battery) • Charging time: ≥ 8 hours (when charged with AC power supply with the power turned off) Power consumption 28VA Classi cation Class I equipment and internally powered equipment, type CF applied part, Continuous Operation, IP22 (Vertically dripping water shall have no harmful effect when the enclosure is tilted at an angle up to 15° from its normal position.) Dimensions Without external communication (RS-232C): 253 mm (W)*1 x 120 mm (H)*1 x 102 mm (D)*2 With external communication (RS-232C): 253 mm (W)*1 x 120 mm (H)*1 x 121 mm (D)*2 *1: Excluding protrusions, *2: Excluding protrusions, pole clamp and moving range